Procleix Panther System Featuring ART

he Procleix Panther system featuring ART provides enhanced levels of automation for laboratories in a compact footprint without compromising throughput. It combines the proven and reliable performance of the Procleix Panther system with software and hardware improvements, to build upon the preferred* NAT screening solution. Innovative optional components, such as full track connectivity and remote monitoring, can be installed to optimize variable workflow needs now or in the future.

*More donations are tested every year with a Procleix system than any other. Grifols, data on file, June 2020.

 

Technical Specifications
Reference Description Presentation
Physical dimensions (WxDxH)
  • Instrument: 122 x 81.5 x 175 cm
  • Instrument with MTU Expansion Module: 149 x 81.5 x 175 cm
*  UPS: 21.4 x 41 x 32.5 cm

  • Direct-to-drain pump: 24 x 22 x 23 cm
Minimum clearance requirements
  • Base: 183 x 172.9 x 198 cm
  • With MTU Expansion Module: 214 x 172.9 x 198 cm
  • No need for rear clearance, instrument can be installed directly against a wall
Weight
  • Base: 363 kg
  • UPS: 34.5 kg
  • With MTU Expansion Module: 423 kg
Sample capacity
  • Up to 120 sample tubes (8 racks) + on-demand access*
Reagent capacity
  • Up to 2000 tests (4 sets) + on-demand access*
Fluids capacity
  • Up to 2000 tests (2 sets) + on-demand access*
Consumables capacity
  • MTU: Up to 125 tests + on-demand access*
  • Tips: Up to 576
  • MTU (with MTU Expansion Module): Up to 600 tests + on-demand access*
Liquid waste capacity
  • Base model: Up to 750 tests + on-demand access*
  • With optional direct-to-drain: No limit**
Solid waste capacity
  • Up to 750 tests+ on-demand access*
Sample throughput
  • 3.5 hr to first 5 results; 5 results every 5 min thereafter
  • 275 test results in 8 hr; up to 500 test results in 12 hr
Sample barcode types
  • Code 39, Code 93, Code 128 (ISBT 128), Interleaved 2 of 5, Codabar, Jan 13, UPC
System calibration
  • Up to 24 hours
On-board reagent stability
  • Up to 60 (also non-continuous) hours
Environmental requirements
  • Altitude: Equal to or less than 2000 m above sea level
  • Operational temperature: 15–30°C
  • Operational humidity: 20–85% relative (non-condensing) at 30°C or below
Electrical requirements
  • Electrical input: 100–240; VAC: (± 10%), 50/60 Hz: single phase
  • Current input: 110 V – 20 amp circuit (dedicated if used with optional UPS) specified to operate at a minimum of 15 amps (dedicated); 220 V – 10 amp circuit (dedicated) specified to operate at 7.5 amps (dedicated)
  • Current draw: Draw 1400 W peak, average 700 W (13 amps @ 100 VAC, 5.4 amps @ 240 VAC)
Environmental output
  • Heat dissipation: 550 W (1878 BTU/hr) during steady state
  • Drain output: Waste fluids drain to removable containers, or directly to centralized laboratory drainage system**
  • Noise output: <60 dB(A)
Operating system
  • Microsoft Windows 10
MTU Expansion Module (optional)
  • Up to 120 MTU capacity (up to 600 tests)
Data sharing (optional)
  • Data sharing among up to 16 connected instruments
  • Exchange partially utilized reagent kits among connected instruments
  • Single user login and setup across networked instruments
  • Duplicate-sample prevention across networked instruments
Dashboard (optional)
  • Provides real-time status of networked Procleix Panther systems featuring ART that can be displayed via TV screen monitor, tablet or desktop PC***

 

 *Allows operator access during system operation; certain restrictions apply
 ** For instruments equipped with optional direct-to-drain pump. Please refer to your local environmental regulations and guidelines
*** Up to 16 Procleix Panther systems featuring ART can be connected, 6 can be displayed at a time